Optimization of the optical particle counter for online particle measurement in high-pressure gas
摘要:With the aim to achieve fast and accurate online measurement of particle size and concentration in different high-pressure gas, the application of an optical particle counter (OPC) is extended from atmospheric pressure to high-pressure conditions (more than 2 MPa). But with the increase of the gas pressure, the performance of the OPC is reduced, resulting in the error of the tested particle size. In order to resolve this problem, an optimization model of optical window thickness is established for the optical sensor of the OPC based on the geometric optics theory. The effects of gas pressure, temperature, and medium on the optimized model are analyzed, which are helpful to improve the accuracy of the optical sensor. Finally, the feasibility, reliability, and accuracy of the method are verified through experiments. (C) 2019 Optical Society of America
关键字:SCATTERING SIZE ANALYSIS MASS CALIBRATION AEROSOLS SHAPE
ISSN号:1559-128X
卷、期、页:卷: 58 期: 2 页: 308-316
发表日期:2019-01-10
影响因子:1.650000
期刊分区(SCI为中科院分区):三区
收录情况:SCI(科学引文索引印刷版),ESI(基本科学指标数据库),EI(工程索引),SCIE(科学引文索引网络版)
发表期刊名称:Applied Optics
通讯作者:卢利锋,王明星
第一作者:吴小林,姬忠礼,熊至宜
论文类型:期刊论文
论文概要:卢利锋,吴小林,姬忠礼,熊至宜,王明星,Optimization of the optical particle counter for online particle measurement in high-pressure gas,Applied Optics,2019,卷: 58 期: 2 页: 308-316
论文题目:Optimization of the optical particle counter for online particle measurement in high-pressure gas